➤ Where several infusion lines are connected on one single vascular access the possibility of their exerting a mutual influence vice-versa cannot be excluded. as caused by change of level) can affect the accuracy of the device. Other components ➤ Variations in pressure (e.g. ➤ If staff call is used we recommend to check the equipment once after connecting the pump. ➤ Do not use the device when service indicator is displayed permanently. Start infusion only if values are corresponding. ➤ Compare displayed value with entered value. ➤ Possible explosion hazard if used in presence of flammable anaesthetics. Observe national specifications and deviations. ➤ Installation in medically used rooms must comply with the appropriate regulations (e.g. ➤ Recommended change of disposables after 24 h (consider national hygiene regulations). ➤ Position the infusion line free of kinks. ➤ Select syringe/catheter suitable for use with the intended medical application. Interrupt the connection during syringe changes to prevent incorrect dose delivery. ➤ Connect to patient only after switching on the device. Also check the device for possible damage. ➤ Prior to use check audible and visual alarms during self test. Operation ➤ Ensure the unit is properly positioned and secured. Application only under regularly supervision by specially trained staff. Patient Safety Attention: Consult accompanying documents !